Technote #21 Where is the emissivity adjustment?

The normal setup and calibration as shown in the

IRt/c Operating Manual and Tech Note #1 automati-

cally compensates for the emissivity and reflectivity

of the material whose temperature is being control-

led, and completely accounts for these effects at

the controlled temperature.

However, for processes in which the control

temperature set-point varies, the control device will

provide higher accuracy over a wider range if its

SPAN or GAIN adjustment is used to calibrate the

process.

Accordingly, the calibration set-up should include a

standard two point method of setting the span.

1. Install IRt/c as close as possible to surface

being measured..

2. Wire connections in standard fashion.

3. At low operating temperature, measure actual

temperature with D Series.

4.  Adjust OFFSET, ZERO, or LO CAL to match reading-

on D Series.

5. At high operating temperature, measure actual

temperature with D Series.

6. Adjust SPAN, GAIN, or HI CAL to match reading

on D Series.